Article Data

  • Views 709
  • Dowloads 138

Original Research

Open Access

Accuracy of an Electronic Apex Locator: A Clinical Evaluation in Primary Molars with and without Resorption

  • Mesut Enes Odabaş1
  • Haluk Bodu1
  • Özlem Tulunoğlu1,*,
  • Alev Alaç am1

1University of Gazi, Faculty of Dentistry, Department of Pediatric Dentistry, 8. Cadde 82.Sokak 06510, Emek Ankara, TURKEY.

DOI: 10.17796/jcpd.35.3.7w71656740286960 Vol.35,Issue 3,May 2011 pp.255-258

Published: 01 May 2011

*Corresponding Author(s): Özlem Tulunoğlu E-mail: ozlemt@gazi.edu.tr

Abstract

The aim of this study was to evaluate the accuracy of electronic apex locator (Root ZX) in primary teeth with and without apical root resorption in vivo. Study Design: The electronic working length was determined in 28 primary teeth (A total of 46 primary molar root canals with 24 roots without resorption and 22 roots with resorption) from twenty-eight healthy children using Root ZX. The teeth were then extracted. The actual length of each root was measured placing a K-File until the tip was visible at the apical foramen or the apical resorption level and approved with digital microscopic evaluation under 20× magnification and then withdrawn 1 mm. The distances from the file tip to the base of the rubber stop were measured and compared with the measurement obtained from apex locator. The data were evaluated with t test using standard statistical software (SPSS 13.0). The critical value of statistical significance was 5%. Results: For root canals without resorption, the accuracy of Root ZX was 79.16% and 95.82% within ± 0.5 and ± 1 mm, respectively. For root canals with resorption, the accuracy of Root ZX was 63.63% and 86.35% within ± 0.5 and ± 1 mm, respectively. No statistically significant differences were detect between the electronically determined lengths and the actual lengths regardless of whether or not resorption. Conclusion: Using a Root ZX Electronic Apex Locator in primary teeth with or without resorption is recommended for the determination of working length.

Keywords

primary teeth, resorption, working length determination, Root ZX, electronic apex locators

Cite and Share

Mesut Enes Odabaş,Haluk Bodu,Özlem Tulunoğlu,Alev Alaç am. Accuracy of an Electronic Apex Locator: A Clinical Evaluation in Primary Molars with and without Resorption. Journal of Clinical Pediatric Dentistry. 2011. 35(3);255-258.

References

1. Kielbassa AM, Muller U, Munz I, Monting JS. Clinical evaluation of the measuring accuracy of ROOT ZX in primary teeth. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 95: 94–100, 2003.

2. Sasaki T, Shimizu T, Watanabe C, Hiyoshi Y. Cellular roles in physiological root resorption of deciduous teeth in the cat. J Dent Res, 69: 67–74, 1990.

3. Sahara N, Toyoki A, Ashizawa Y, Deguchi T, Suzuki K. Cytodifferentiation of the odontoclast prior to the shedding of human deciduous teeth: an ultrastructural and cytochemical study. Anat Rec, 244: 33–49. 1996.

4. Rodd HD, Waterhouse PJ, Fuks AB, Fayle SA, Moffat MA. Pulp therapy for primary molars. Int J Paediatr Dent, 16 Supp l 1: 15–23, 2006.

5. Olson AK, Goerig AC, Cavataio RE, Luciano J. The ability of the radiograph to determine the location of the apical foramen. Int Endod J, 24: 28–35, 1991.

6. Tamse A, Kaffe I, Fishel D. Zygomatic arch interference with correct radiographic diagnosis in maxillary molar endodontics. Oral Surg Oral Med Oral Pathol, 50: 563–6, 1980.

7. Suzuki K, Sunada I, Nagasawa W. A study of the technique of root canal theraphy. J Dent Res, 41: 58, 1962.

8. Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electronic apex locator that uses the ratio method in vital and necrotic canals. J Endod, 24: 48–50, 1998.

9. ElAyouti A, Weiger R, Lost C. The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length. J Endod, 28: 116–9, 2002.

10. Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod, 19: 177–9, 1993.

11. Fouad AF, Krell KV, McKendry DJ, Koorbusch GF, Olson RA. Clinical evaluation of five electronic root canal length measuring instruments. J Endod, 16: 446–9, 1990.

12. Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod, 22: 616–8, 1996.

13. Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator. J Endod, 24: 438–41, 1998.

14. Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod, 20: 111–4, 1994.

15. Berman LH, Fleischman SB. Evaluation of the accuracy of the Neosono-D electronic apex locator. J Endod, 10: 164–7, 1984.

16. Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variations in canal irrigants and foramen size. J Endod, 19: 63–7, 1993.

17. Saito T, Yamashita Y. Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants. Dent Jpn (Tokyo), 27: 65–72, 1990.

18. Mente J, Seidel J, Buchalla W, Koch MJ. Electronic determination of root canal length in primary teeth with and without root resorption. Int Endod J, 35: 447–52, 2002.

19. Bodur H, Odabas M, Tulunoglu O, Tinaz AC. Accuracy of two different apex locators in primary teeth with and without root resorption. Clin Oral Investig, 12: 137–41, 2008.

20. Katz A, Mass E, Kaufman AY. Electronic apex locator: a useful tool for root canal treatment in the primary dentition. ASDC J Dent Child, 63: 414–7, 1996.

21. Leonardo MR, Silva LA, Nelson-Filho P, Silva RA, Raffaini MS. Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth. Int Endod J, 41: 317–21, 2008.

22. Tosun G, Erdemir A, Eldeniz AU, Sermet U, Sener Y. Accuracy of two electronic apex locators in primary teeth with and without apical resorption: a laboratory study. Int Endod J, 41: 436–41, 2008.

23. Angwaravong O, Panitvisai P. Accuracy of an electronic apex locator in primary teeth with root resorption. Int Endod J, 42: 115–21, 2009.

24. ElAyouti A, Kimionis I, Chu AL, Lost C. Determining the apical terminus of root-end resected teeth using three modern apex locators: a comparative ex vivo study. Int Endod J, 38: 827–33, 2005.

25. Nguyen HQ, Kaufman AY, Komorowski RC, Friedman S. Electronic length measurement using small and large files in enlarged canals. Int Endod J, 29: 359–64, 1996.

26. Meares WA, Steiman HR. The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator. J Endod, 28: 595–8, 2002.

27. Pommer O, Stamm O, Attin T. Influence of the canal contents on the electrical assisted determination of the length of root canals. J Endod, 28: 83–5, 2002.

28. Ghaemmaghami S, Eberle J, Duperon D. Evaluation of the Root ZX apex locator in primary teeth. Pediatr Dent, 30: 496–8, 2008.

29. Brunton PA, Abdeen D, MacFarlane TV. The effect of an apex locator on exposure to radiation during endodontic therapy. J Endod, 28: 524–6, 2002.

Abstracted / indexed in

Science Citation Index Expanded (SciSearch) Created as SCI in 1964, Science Citation Index Expanded now indexes over 9,500 of the world’s most impactful journals across 178 scientific disciplines. More than 53 million records and 1.18 billion cited references date back from 1900 to present.

Biological Abstracts Easily discover critical journal coverage of the life sciences with Biological Abstracts, produced by the Web of Science Group, with topics ranging from botany to microbiology to pharmacology. Including BIOSIS indexing and MeSH terms, specialized indexing in Biological Abstracts helps you to discover more accurate, context-sensitive results.

Google Scholar Google Scholar is a freely accessible web search engine that indexes the full text or metadata of scholarly literature across an array of publishing formats and disciplines.

JournalSeek Genamics JournalSeek is the largest completely categorized database of freely available journal information available on the internet. The database presently contains 39226 titles. Journal information includes the description (aims and scope), journal abbreviation, journal homepage link, subject category and ISSN.

Current Contents - Clinical Medicine Current Contents - Clinical Medicine provides easy access to complete tables of contents, abstracts, bibliographic information and all other significant items in recently published issues from over 1,000 leading journals in clinical medicine.

BIOSIS Previews BIOSIS Previews is an English-language, bibliographic database service, with abstracts and citation indexing. It is part of Clarivate Analytics Web of Science suite. BIOSIS Previews indexes data from 1926 to the present.

Journal Citation Reports/Science Edition Journal Citation Reports/Science Edition aims to evaluate a journal’s value from multiple perspectives including the journal impact factor, descriptive data about a journal’s open access content as well as contributing authors, and provide readers a transparent and publisher-neutral data & statistics information about the journal.

Scopus: CiteScore 2.0 (2022) Scopus is Elsevier's abstract and citation database launched in 2004. Scopus covers nearly 36,377 titles (22,794 active titles and 13,583 Inactive titles) from approximately 11,678 publishers, of which 34,346 are peer-reviewed journals in top-level subject fields: life sciences, social sciences, physical sciences and health sciences.

Submission Turnaround Time

Conferences

Top